You are here

Low-cost High-resolution fundus camera

Award Information
Agency: Department of Health and Human Services
Branch: National Institutes of Health
Contract: 2R44EY018071-02
Agency Tracking Number: EY018071
Amount: $778,218.00
Phase: Phase II
Program: SBIR
Solicitation Topic Code: N/A
Solicitation Number: PHS2007-2
Timeline
Solicitation Year: 2008
Award Year: 2008
Award Start Date (Proposal Award Date): N/A
Award End Date (Contract End Date): N/A
Small Business Information
2501 Yale Blvd. SE Suite 201
ALBUQUERQUE, NM 87106
United States
DUNS: 183651723
HUBZone Owned: No
Woman Owned: No
Socially and Economically Disadvantaged: Yes
Principal Investigator
 PETER SOLIZ
 (505) 228-3875
 PSOLIZ@VISIONQUEST-BIO.COM
Business Contact
Phone: (505) 798-2505
Email: psoliz@visionquest-bio.com
Research Institution
N/A
Abstract

DESCRIPTION (provided by applicant): The principal objective is to demonstrate clinically a commercially viable, low-cost, high-resolution fundus camera that: is easy to use, i.e., consistent with the operation of current fundus cameras; has a field of vie
w (FOV) that is typical of current fundus cameras (15 to 30 degrees); has a factor of five or more improvement in resolution as compared to existing clinical instruments; and, will be priced below that of today's clinical devices. In Phase I, a low-cost me
ans for removing low-order aberrations, focus and astigmatism, was demonstrated. It was shown that removal of low-order aberrations would result in significantly better resolution images. In Phase II, the specific aims are to integrate the low-cost low-ord
er aberrations compensator into a scanning laser ophthalmoscope (SLO). This new SLO would produce retinal and optic disc images that have five times better resolution when compared to today's commercial cameras. Researchers have sought to gain greater insi
ght into the mechanisms of the retina and the optic disc at high spatial resolutions that would enable the visualization of structures (4 to 8 micrometers) such as capillaries and nerve fiber bundles. Major sources of retinal image quality degradation are
aberrations within the human eye. The presence of these ocular aberrations, many of them random and non deterministic, limits the achievable resolution and the contrast of small image details due to diffraction effects. To overcome these fundamental limita
tions, researchers have been applying adaptive optics techniques to correct for the aberrations. Today, deformable mirror based adaptive optics devices have been developed to overcome the limitations of standard fundus cameras, but at prices in excess of
500,000. The device proposed by VisionQuest addresses the same goals as the above research grade adaptive optics instruments but does so in a manner which will ultimately be better suited for clinical use because of its similarity to current retinal photog
raphic techniques and its lower cost technology. Our proposed device avoids the high costs, complexity, and limited FOV by adopting unique and notably less expensive wavefront measurement and aberration correction techniques. VisionQuest's goal is to demon
strate an advanced imaging technology that introduces an effective and low cost ( 10,000 or less) diagnostic tool to the healthcare system. PUBLIC HEALTH RELEVANCE: Retinal imaging for the purposes of diagnosis of retinal diseases or for monitoring the tre
atment and/or progression of retinal diseases is standard of care. Greatly improved images of the retina, especially at costs that are lower than current fundus cameras, would have a significant impact on our ability to implement broad-scale screening thro
ughout the U.S. for retinal disease. A low-cost retinal camera would enable the early detection and care of individuals with any of the three major eye diseases, i.e., age-related macular degeneration, diabetic retinopathy, or glaucoma.

* Information listed above is at the time of submission. *

US Flag An Official Website of the United States Government